LVAD Outflow Cannula Occlusion with Thrombus and Native Aortic Valve Opening
▶ Watch on YouTube — age-restricted, cannot embed here
Imaging findings
Gated cardiac CT demonstrates complete thrombotic occlusion of the Left Ventricular Assist Device (LVAD) outflow cannula, with thrombus propagating into the pump itself. Crucially, the native aortic valve is observed opening with every cardiac beat, indicating significant LVAD dysfunction due to the outflow obstruction.
Key takeaways
Complete thrombotic occlusion of an LVAD outflow cannula is a serious complication leading to pump failure. A key indicator of LVAD dysfunction is the opening of the native aortic valve during systole, which should normally remain closed due to the continuous flow provided by a properly functioning LVAD. Gated CT is invaluable for assessing both cannula patency and native valve dynamics.
AI-assisted summary — may contain errors. Verify against the source video. Learn more
More from this webinar
- Atrial Septal Defect (ASD)Congenital
- Variant of Levoatriocardinal VeinCongenital
- Active Intercostal Artery Bleed with HemothoraxTrauma
- Hughes-Stovin SyndromeAutoimmune
- Excipient Lung Disease (Talc Granulomatosis)Exposure
- Lipomatous Hypertrophy of the Interatrial Septum with FDG AvidityArtifact
- Acute Type A Aortic Dissection with Malperfusion and StrokeVascular
- Bilateral Superior Vena Cava with Left SVC to Azygos DrainageCongenital
See all cases from February 26, 2016 →
Related Iatrogenic cases
- Embolized Amplatzer Vascular Plugs Causing Claudication
- CoreValve Dislodgement and Fixation with Palmaz Stent During TEVAR
- Dislodged Tracheoesophageal Puncture Speech Prosthesis
- Post-Transplant Lung Torsion
- Recurrent Mediastinal Seroma Post Aortic Repair
- Iatrogenic Hemothorax from Medially Placed Chest Tube Injury to Intercostal Artery