Riata Pacemaker Lead Externalization

Imaging findings

Radiograph and lateral view of a pacemaker system, when closely magnified, reveal structural failure with externalization of the inner conductor cables from the main lead body.

Key takeaways

Riata pacemaker leads are prone to a specific failure mechanism known as externalization, where the inner conductors wear through the outer silicone insulation. High-resolution magnification on radiographs is often required to detect this defect, which necessitates lead replacement due to the risk of electrical failure or inappropriate shocks.