Aortic Flow mixing (Smoke) Artifact

Imaging findings

Contrast-enhanced CT angiogram shows a prominent, horizontal, low-attenuation band of poor contrast mixing (or "smoke" artifact) in the transverse aortic arch and proximal descending thoracic aorta. A repeat CT angiogram obtained with a slightly delayed arterial phase demonstrates complete resolution of the low-attenuation band with normal, homogeneous opacification of the aorta.

Key takeaways

Slower blood flow along the inner curvature (undersurface) of the aortic arch can cause transient contrast-mixing artifacts that mimic acute aortic dissection on early-phase, rapid-pitch helical acquisitions. This phenomenon of slower flow also explains why aortic thrombi preferentially develop in this specific location.