LVAD Outflow Cannula Extraluminal Fluid Mimicking Thrombus
Imaging findings
In a patient with a longstanding left ventricular assist device presenting with new low-flow alarms, CT of the outflow cannula shows near-total luminal collapse with an unusual star- or propeller-shaped configuration, distinct from the more typical eccentric appearance of true thrombus.
Key takeaways
This appearance represented fluid accumulated within the graft lining between the outflow cannula and its outer sheath rather than true thrombus; drainage of the fluid immediately improved flow, and the atypical propeller/star shape -- rather than an eccentric thrombus-like mass effect -- is a useful clue that a low-flow LVAD alarm reflects extraluminal fluid rather than intraluminal thrombus.
AI-assisted summary — may contain errors. Verify against the source video. Learn more
More from this webinar
- Radiation Recall PneumonitisILD
- Disseminated Candidiasis with Miliary Pulmonary NodulesInfection
- Marijuana-Associated Aspergillus BronchitisInfection
- SVC Syndrome with Venous-Enhancement Pseudo-Sclerotic Bone MetastasesArtifact
- Disseminated Coccidioidomycosis Mimicking Metastatic CancerInfection
- LVAD Driveline Infection Extending to the Outflow CannulaInfection
- Fatal Reexpansion Pulmonary Edema with Contralateral ARDSInflammatory
- Necrotizing Sarcoid Granulomatosis with CNS and Pulmonary InvolvementAutoimmune
See all cases from July 19, 2018 →
Related Artifact cases
- Tipped-Up Right Middle Lobe (Chronic Cicatrization Atelectasis)
- Post-Lobectomy Mediastinal Lymphocele Mimicking Recurrent Lymphadenopathy
- Interrupted Contrast Bolus Artifact Mimicking Pulmonary Embolism
- Bioabsorbable Hemostatic Packing Material (Surgicel) Mimicking a Mass After Aortic Surgery
- Prominent Right Third Intercostal-Bronchial Artery Trunk Mimicking Aortic Injury
- Slow-Flow Mixing Artifact (Pseudo-Thrombus) in the Aorta and Left Atrial Appendage