Externalization of Pacemaker Lead (Medtronic Riata Lead)
Imaging findings
On chest radiograph, a subtle displacement of the conducting electrode away from the right ventricular lead is observed, indicating externalization. Cine fluoroscopy during the procedure provided better visualization of this displacement. A photograph of the extracted lead clearly shows the blue conductor completely outside its insulation. Externalization of a coronary vein lead was also subtly suggested.
Key takeaways
Externalization is a notorious mode of failure for certain pacemaker leads, specifically the Medtronic Riata and St. Jude leads, where insulation breakdown exposes the conducting electrodes. It presents as a subtle but critical finding on chest radiographs, which clinicians must recognize to prevent serious complications. The cardiologist specifically looked for this problem based on the known lead type.
AI-assisted summary — may contain errors. Verify against the source video. Learn more
More from this webinar
- SBRT Radiation PneumonitisILD
- Loeys-Dietz Syndrome with Vessel TortuosityCongenital
- Delayed Presentation of Myocardial Infarction with Left Ventricular Thrombus and Dressler's SyndromeVascular
- Excipient Lung DiseaseExposure
- Rheumatoid Arthritis-Associated Cystic Lung DiseaseILD
- Transposition of Great Arteries Post-Arterial Switch with Lecompte Maneuver and Main Pulmonary Artery CompressionCongenital
- Right Atrial Myxoma with Atypical MRI CharacteristicsNeoplastic
- Paraneoplastic Pemphigus with Castleman's Disease and Constrictive BronchiolitisAutoimmune
See all cases from June 11, 2016 →
Related Iatrogenic cases
- Right Ventricular Lead Perforation
- Aortic Annulus Rupture during TAVI
- Impella Device Fracture with Free Fragment in Left Ventricle
- Phrenic Nerve Injury (Left Hemidiaphragm Paralysis) with Crus Atrophy secondary to Spinal Cord Emboli during Aortic Stent Graft Placement
- Infected Breast Reduction Seroma / Abscess
- Cement Embolization to Pulmonary Arteries after Vertebroplasty