180-Degree Flip of a Defibrillator along the Long Axis
▶ Watch on YouTube — age-restricted, cannot embed here
Imaging findings
Serial chest radiographs demonstrate a defibrillator generator that has rotated 180 degrees along its long axis within the subcutaneous pocket. On the baseline radiograph, the lettering on the generator reads normally, whereas the subsequent radiograph shows the lettering as a mirror image, indicating a complete flip of the device. The leads remain intact without evidence of coiling, tension, or dislodgement.
Key takeaways
Unlike classic Twiddler's syndrome, which involves short-axis rotation and lead coiling, simple flipping of a device along its long axis can occur within a capacious subcutaneous pocket. If the device is interrogated and found to be functioning normally with adequate capture and lead integrity, surgical revision may not be necessary.
AI-assisted summary — may contain errors. Verify against the source video. Learn more
More from this webinar
- Patent Ductus Arteriosus (PDA) and Necrotizing Enterocolitis (NEC)Congenital
- Daptomycin-Induced Pulmonary ToxicityILD
- Strongyloides Hyperinfection SyndromeInfection
- Polymyositis-Associated Interstitial Lung Disease (NSIP / Organizing Pneumonia Overlap)ILD
- Cytomegalovirus (CMV) Pneumonia with Organizing PneumoniaInfection
- Typical Type A Acute Aortic DissectionVascular
- Type A Aortic Dissection with Retrograde Valve Prolapse and Intimal IntussusceptionVascular
- Atrial Extension of Mitral Annular Calcification (MAC)Metabolic
See all cases from November 23, 2014 →
Related Iatrogenic cases
- Barium Aspiration during Swallow Study
- Buffalo Chest (Bilateral Pneumothoraces with Pleural Communication)
- Aortoesophageal Fistula from Nasogastric Intubation with Aberrant Aortic Arch
- Iatrogenic Esophageal Rupture Post-Intubation
- Post-Lobectomy Unilateral Pulmonary Edema from Kinked Pulmonary Artery
- Lung Torsion Post-Upper Lobectomy