CTICases ← All cases

Corduroy Artifact in Digital Radiography

Imaging findings

Chest radiographs demonstrate faint, regular vertical striations (resembling grid lines) across the image plate. This occurred during portable digital radiography in patients with active mechanical devices (an Impella pump in one case, and a HeartMate LVAD in another).

Key takeaways

The 'corduroy' or electromagnetic interference artifact in digital radiography is caused by radiofrequency or electromagnetic emissions from the spinning motors of left ventricular assist devices (LVADs) or Impella pumps. This interference affects the digital plate's electronics during readout, producing characteristic grid-like lines.

AI-assisted summary — may contain errors. Verify against the source video. Learn more

More from this webinar

See all cases from February 12, 2022 →

Related Artifact cases

Browse all Artifact cases →